Notes:
1.Before operation, the power must be disconnected, otherwise there is a risk of electric shock.
2.Please keep the screws that are removed properly to avoid losing them.
Hex wrench,flathead screwdriver,Phillips screwdriver,door cover key
①.Remove the water pipe from the outlet, wait for the water in the glass pipe to flow back into the chiller, and then reconnect the water pipe from the outlet
②.Open the upper rear door and the lower left rear door of the equipment.
③.Use a steel ruler to measure the distance from the glass tube to Mirror 1 and record it.
④.Disconnect the positive and negative electrode wires:
⑤.Cut the zip ties with a utility knife and remove the inlet and outlet water pipes from the glass tube.
⑥.Use a hex wrench to remove the screws securing the glass tube and then take out the glass tube.
①.Place the laser on the laser holder and adjust the distance from the laser to Mirror 1 (based on the distance recorded during disassembly).
②.Tighten the tape while securing the laser with screws.
③.Connect the inlet and outlet water pipes to the laser and secure them with zip ties.
④.Reconnect the positive and negative output wires.
①.Open the lower left side door of the equipment.
②.Disconnect the power cable, fan power cable, and signal cable.
③.Rotate counterclockwise to unscrew the beam expander, and then use a hex wrench to remove the beam expander mount
④.Remove the four fixing screws of the RF tube and take out the RF tube.
⑤.Use a hex wrench to separate the RF tube from the RF tube holder.
①.Install the beam expander mirror holder onto the new RF tube, and then install the beam expander mirror onto the mirror holder
②.Attach the RF tube holder to the new RF tube.
③.Secure the RF tube inside the machine.
④.Reconnect the RF tube power cable, fan power cable, and communication network cable.