Laser replacement

Laser replacement

Notes:

1.Before operation, the power must be disconnected, otherwise there is a risk of electric shock.

2.Please keep the screws that are removed properly to avoid losing them.

Tools Required:

Hex wrench,flathead screwdriver,Phillips screwdriver,door cover key

Glass Tube Replacement

1.Disassembly of Glass Tube

.Remove the water pipe from the outlet, wait for the water in the glass pipe to flow back into the chiller, and then reconnect the water pipe from the outlet

②.Open the upper rear door and the lower left rear door of the equipment.

 

③.Use a steel ruler to measure the distance from the glass tube to Mirror 1 and record it.

④.Disconnect the positive and negative electrode wires:

  1. Rotate the positive electrode terminal counterclockwise.

  1. Use a flathead screwdriver to loosen the screw securing the negative electrode wire.

⑤.Cut the zip ties with a utility knife and remove the inlet and outlet water pipes from the glass tube.

⑥.Use a hex wrench to remove the screws securing the glass tube and then take out the glass tube.

2.Installation of Glass Tube

①.Place the laser on the laser holder and adjust the distance from the laser to Mirror 1 (based on the distance recorded during disassembly).

②.Tighten the tape while securing the laser with screws.

③.Connect the inlet and outlet water pipes to the laser and secure them with zip ties.

④.Reconnect the positive and negative output wires.

RF Tube Replacement

1.Disassembly of RF Tube

①.Open the lower left side door of the equipment.

 

②.Disconnect the power cable, fan power cable, and signal cable.

③.Rotate counterclockwise to unscrew the beam expander, and then use a hex wrench to remove the beam expander mount

④.Remove the four fixing screws of the RF tube and take out the RF tube.

⑤.Use a hex wrench to separate the RF tube from the RF tube holder.

2.Installation of RF Tube

①.Install the beam expander mirror holder onto the new RF tube, and then install the beam expander mirror onto the mirror holder

②.Attach the RF tube holder to the new RF tube.

③.Secure the RF tube inside the machine.

④.Reconnect the RF tube power cable, fan power cable, and communication network cable.