Note: To ensure the normal operation of the One Laser XT model, please note that the cold water pump equipped with this device is used to cool the glass tube. The water in the water tank of the water pump has been drained at the factory. If it is powered on directly without adding pure water, it will directly cause electrical failure. Therefore, after receiving the device, please add pure water to the water tank of the water pump as soon as possible to ensure the safe operation of the device.
1.Preparation
①. Consumable preparation
One Laser XT model device*1,M2 hexagon wrench
②. Debug software
None
2.Operation step
①.Remove the back cover of the device
Note:The left door of the device needs to be removed at the same time.
- Use an M2 hexagonal wrench to remove the hexagonal screws on the back cover and left door of the device and place them together.
- Take out the back cover of the device
②.Fill the water tank with pure water
Note: Using pure water as cooling water can prevent scaling and corrosion, improve cooling efficiency, and inhibit the growth of microorganisms, thereby effectively protecting the equipment, ensuring its stable operation and extending its service life.
- Find the location of the water tank
- Open the water filling cap on the top of the water tank and pour in about 790ml of pure water.
Older models;
- Start the equipment, after the pump is started, immediately pour in about 790ml of pure water, observe while pouring, until the water in the storage tank flows into the glass tube .
- After has flowed into the tube, you need to add another 150ml or so of pure water, and once you have done so, put the filler cap back on and tighten it.
- Find the location of the water tank